Factory Outlets Finger Blood Oxygen - Patient Monitor P9000N – Medke Detail:
P9000N Multi-Para Patient monitor
Features &Benefits:
5 standard parameters: ECG, RESP, NIBP, SPO2, 1-TEMP.
12.1”high resolution color TFT display..
Real time S-T segment analysis, pace-maker detection and ARR analysis.
Multi-display selectable including standard, large font, trend coexis, OxyCRG dynamic
On-line help and Patient info input management
Multi-lead ECG waveforms display in phase.
Large volume of tabular and graphic trends information storage and easy to recall
Capture dynamic waveforms.
Efficient resistance to interference of defibrillator and HF knife.
Up to 4 hours working capacity of built-in rechargeable battery.
Networking capacity and TCP/IP networking platform.
Option of FHR,FM,TOCO,Printer , IBP and EtCO2
Performance Specifications
Display: 12.1’’color TFT
Rolling and refreshing waveform display
Multi display selectable including:
Large-font display
Trend coexist display
OxyCRG dynamic view display.
Bed-to-bed view display
Trace: 9 waveforms(7 ECG, 1 SPO2 and 1 RESP)
Sweep speed:12.5mm/s,25mm/s,50mm/s
Indicator: Power/battery indicator light
QRS beep and alarm sound
Battery: Rechargeable lead acid cell, 12v/4AH
Max.24 hour for charging, 4 hours for continue working
Trend: Parameter graphic and tabular trends:
5s/piece,8 hours;
1min/piece,168 hours(24hour×7days)
5min/piece,1000 hours.
Storage: NIBP:1000 groups
Alarm:200 groups
Full disclosure waveforms:1 hour
Alarm: User-adjustable High, Medium and Low limits 3-level Audible and visual alarm
Networking: Connected to central monitoring system
TCP/IP netting platform
Standard Parameters
ECG:
Lead mode: 5 –lead(R,L,F,N,C)
Lead selection : I,II,III,avR,avL,avF,V
Waveform: 3 and 7 channel selectable
Gain selection: 0.5mm/mv,1mm/mv,2mm/mv
Sweep speed: 12.5mm/s;25mm/s;50mm/s
Heart rate range:
Adult: 15~300bpm;
Neonate:/pediatric:15~350bpm
Accuracy: +1bpm or +1%,whichecer is greater
Resolution: 1bpm
Filter: surgery mode:1~20Hz
monitor model:0.5~40Hz
Diagnostic mode:0.05~130Hz
Scaling signal: 1mv,+3%
Protection: withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation
Alarm range: 15~350bpm
S-T segment detecton:
Measurement range: 2.0mV~+2.0mV
Alarm Range: -2.0mV~ +2.0mV
Accuracy: -0.8mV ~+0.8Mv
Error: +0.02Mv
Arrhythmia analysis: YES
SPO2
Measurement range: 0~100%
Resolution: 1%
Accuracy: +2%(70-100%);0-69% unspecified
Alarm range 0~100%
Pluse Rate: range:20~300bpm
Resolution:1bpm
Error:+1bpm or +2%,whichever is greater
NIBP
Method: Digital Automatic oscillometric
Operation mode : Manual/Automatic/continuous
Auto measurement time: Adjustable(1~480min)
Measurement Unit: mmHg/Kpa selectable
Measurement types: Systolic, Diastolic, Mean
Measurement tange:
Range of Systolic pressure:
Adult: 40~270mmHg
Pediatric: 40~220mmHg
Neonate: 40~135mmHg
Range of mean pressure:
Adult: 20~235mmHg
Pediatric: 20~165mmHg
Neonate: 20~110mmHg
Range of diastolic pressure:
Adult: 10~215mmHg
Pediatric: 10~150mmHg
Neonate :10~100mmHg
Over-pressure protection:
Double safety protection
Resolution: 1 mmHg
Alarm: Systolic.Diastolic,Mean
RESPIRATION
Method: Thoracic impedance
Measurement range: Adult:7~120rpm;
Neonate/Pediatric:7~150rpm
Apnea alarm: YES,10~40s
Resolution: 1rpm
Accuracy: +2rpm
TEMPERATURE
Compatible probe: YSI or CYF
Measurement range: 5~50℃
Resolution : 0.1℃
Accuracy: +0.1℃
Refreshing time: about 1
Average measuring time: <10s
FHR
Transducer: Multi-crystal, Pulsed Doppler
Measurement range: 50~210 BPM
Working frequency: 1 MHz
Strength: <5mW/cm2
Signal processing:
special DSP system and modern recognization.
Resolution: 1BPM
Accuracy: ±1BPM
Alarm Range: High: 160,170,180,190 BPM,
Low: 100,110,120 BPM
FM
Manual button marking,
automatic FM identifying function
TOCO Measuring
Transducer: External pressure transducer
Measurement range: 0~100 units
Resolution: 1rpm
Accuracy: ±2 rpm
IBP
Channel: 2 channels
Range: -50-300mmHg
Resolution: 1mmHg
Accuracy: ±4mmHg(±4%)
Unit: mmHg, Kpa
Converter Sensitivity: 5mV/V/mmHg
Transducer sites: ART/PA/CVP/LAP/RAP/ICP
EtCO2( Sidertream CO2)
Measurement range: 0~99mmHg
Accuracy: +2mmHg (0~40mmHg)
Sampling Range: 100ml/min
Sampling rate accuracy: 15%
Respiration rate: 0~120rmp
Respiration accuracy: +2rmp (0~70rmp)
+5rmp (>70rmp)
Respiration time: <240msec (10% to 90%)
Delay time: <2s
EtCO2 (mainstreaam CO2)
Method: Infrared Spectrum
Range: 0.0-10% (0~76%)
Resolution: 1mmHg (0.1%)
Accuracy: <5%(±4.0 mmHg)
Or<10%(of Readings)
Recorder:
Build-in, thermal array
Plethysmogram caveform:3 channels
Record mode: manual, on alarm, time-defined
Recording width: 50mm
Printing speed: 50mm/s
Recording type: Frozen waveform record
NIBP recall record
Trend table record
Alarm record
Fixed -time record
Miscellaneous
Satety:
Safety level: Class I, type CF
Dimendion and Weight
Dimendion: 440×430×450
G.Weight: <9.0KS
Operation Environment
Temperature: Working 0~+40℃
Transportation and storage –20~+60℃
Humidity: working≤85%
Transportation and storage≤93%
Power: AC 100-240,50/60Hz
Patient range:
Neonate, pediatric, and adult patients
Standard Accessories:
(1) 5 lead ECG cable
(2) 1 spo2 probes
(3) 1 NIBP prbe
(4) 1 temp probe
(5)1 ground connecting liner
(6) Chest electrode (10pcs/set)
Product detail pictures:
Related Product Guide:
IRADIMED CORP (NASDAQ:IRMD) Stock Price Up as Sentiment Increases | Fingertip Oximeter
Heart attacks: These bands now offer medical grade alerts | Ecg Leadwires
We depend on sturdy technical force and continually create sophisticated technologies to meet the demand of Factory Outlets Finger Blood Oxygen - Patient Monitor P9000N – Medke, The product will supply to all over the world, such as: Las Vegas , Manchester , Greece , Based on products and solutions with high quality, competitive price, and our full range service, we have accumulated experienced strength and experience, and we've built up a very good reputation in the field. Along with the continuous development, we commit ourselves not only to the Chinese domestic business but also the international market. May you moved by our high quality items and passionate service. Let's open a new chapter of mutual benefit and double win.
By Tina from Ottawa - 2015.12.14 15:26
The customer service staff is very patient and has a positive and progressive attitude to our interest, so that we can have a comprehensive understanding of the product and finally we reached an agreement, thanks!
By Esther from Greece - 2015.06.30 17:29